part_subject: "CONTROL OF EMISSIONS FROM NONROAD SPARK-IGNITION ENGINES AT OR BELOW 19 KILOWATTS"
subpart_code: "D"
subpart_subject: "Emission Test Equipment Provisions"
section_number: "90.305"
section_subject: "Dynamometer specifications and calibration accuracy."
cfr_reference: "40 CFR 90.305"
title_name: "Title 40"
title_subject: "Protection of Environment"
parts_covered: "Parts 87 to 95"
revised_date: "Revised as of July 1, 2017"
publication_date: "As of July 1, 2017"
contains_description: "Containing a codification of documents of general applicability and future effect"
publication_info: "Published by the Office of the Federal Register National Archives and Records Administration as a Special Edition of the Federal Register"
---
(a)Dynamometer specifications. The dynamometer test stand and other instruments for measurement of speed and power output must meet the engine speed and torque accuracy requirements shown in Table 2 in Appendix A of this subpart. The dynamometer must be capable of performing the test cycle described in § 90.410.
(b)Dynamometer calibration accuracy. (1) The dynamometer test stand and other instruments for measurement of power output must meet the calibration frequency shown in Table 2 in Appendix A of this subpart.
(2)A minimum of three calibration weights for each range used is required. The weights must be equally spaced and traceable to within 0.5 percent of National Institute for Standards and Testing (NIST) weights. Laboratories located in foreign countries may certify calibration weights to local government bureau standards.