This section describes the equations for calculating molar flow rates from various flow meters. After you calibrate a flow meter according to § 1065.640, use the calculations described in this section to calculate flow during an emission test.
(a)PDP molar flow rate. (1) Based on the speed at which you operate the PDP for a test interval, select the corresponding slope, a
1, and intercept, a
0, as calculated in § 1065.640, to calculate PDP molar flow rate,, as follows:
through one venturi or one combination of venturis, use its respective mean C
d and other constants you determined according to § 1065.640 and calculate n
as follows:
(2)To calculate the molar flow rate through one venturi or a combination of venturis, you may use its respective mean, K
v, and other constants you determined according to § 1065.640 and calculate its molar flow rate n
during an emission test. Note that if you follow the permissible ranges of dilution air dewpoint versus calibration air dewpoint in Table 3 of § 1065.640, you may set M