part_subject: "GENERAL AND PLASTIC SURGERY DEVICES"
subpart_code: "E"
subpart_subject: "Surgical Devices"
section_number: "878.5040"
section_subject: "Suction lipoplasty system."
cfr_reference: "21 CFR 878.5040"
title_name: "Title 21"
title_subject: "Food and Drugs"
parts_covered: "Parts 800 to 1299"
revised_date: "Revised as of April 1, 2019"
publication_date: "As of April 1, 2019"
contains_description: "Containing a codification of documents of general applicability and future effect"
publication_info: "Published by the Office of the Federal Register National Archives and Records Administration as a Special Edition of the Federal Register"
---
(a)Identification. A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
(b)Classification. Class II (special controls). Consensus standards and labeling restrictions.