contains_description: "Containing a codification of documents of general applicability and future effect"
publication_info: "Published by the Office of the Federal Register National Archives and Records Administration as a Special Edition of the Federal Register"
---
(a)Identification. A tympanic membrane contact hearing aid is a prescription device that compensates for impaired hearing. Amplified sound is transmitted by vibrating the tympanic membrane through a transducer that is in direct contact with the tympanic membrane.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)The patient contacting components must be demonstrated to be biocompatible.
(2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, and must include:
(vi)Validation testing of device output and mechanical force applied to the tympanic membrane in a clinically appropriate model.
(3)Clinical performance testing must characterize any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.
(4)Professional training must include the ear impression procedure, correct placement, fitting, monitoring, care, and maintenance of the device.
(A)A patient card that identifies if a patient has been fitted with any non-self- removable components of the device and provides relevant information in cases of emergency;
(B)Information on how to correctly use and maintain the device;
(C)The potential risks and benefits associated with the use of the device; and